Light source
- Quartz-halogen lamp 12V/20W
Wavelength range
- Automatic by 8 positions filter-wheel
- 7 standard filters: 340, 405, 492, 510, 546, 578, 630nm,
- 1 position free for optional filter
- Bandwidth < 8nm
- Stray light: < 1.0 % @ 340 nm
Photometric range
- 0.0000 to 3.0000 ABS
- Resolution : 0.0001 ABS
- Stability : 0.002 A/hour
Flowcell
- Stainless steel with quartz window
- Measuring volume 32 µl
- Optical path 10mm
- Aspiration volume programmable : 200 – 2000 µl
Analysis method
- End point, with or without reagent blank
- End point, with sample blank and with or without reagent blank
- Kinetic, with or without linearity check
- Two point kinetic / fixed time, with or without reagent blank
- Absorbance
- Turbidimetry
- Linear and non-linear calibration
PARAMETER SETTINGS
- Method
- Wavelength
- Temperature
- Reagent blank y/n
- Sample blank y/n
- Lag time
- Measurement time
- Reaction type
- Absorbance limit
- Aspiration volume
- Standards
- Linearity check
- Unit for results
Temperature control
- By means of Peltier elements
- 25°, 30°, 37°C +-0.1°C
Memory
Input
Display
Printer
Interface
Operating environment
- Temperature : 15 – 30 °C
- Humidity : 20 – 80 %
Power requirement
- AC 220V +/- 10%, 50/60Hz
- AC 110V +/- 10%, 50/60Hz
Dimension
Weight